Evaluation of the sterilization efficacy of domestic electric drills used in orthopaedic surgeries

Aún no traducido Aún no traducido
Categoría Estudio primario
RevistaBraz. j. microbiol
Año 2009

Este artículo no está incluido en ninguna revisión sistemática

Cargando información sobre las referencias
It is estimated that electric drills (ED) have been used in orthopaedic surgeries for bone drilling for more than 50 years in Brazilian hospitals. It is an electric, thermosensitive equipment, not indicated for surgical use, which has not been previously evaluated regarding the sterilization efficacy, being suspect of infection risk. This study evaluated the efficacy of sterilization by ethylene oxide (EtO) of new drills that were intentionally contaminated with Bacillus atrophaeus spores. An experimental, laboratory, randomized applied research was developed, where 16 electric drills were analyzed, in addition to positive and negative controls. All the previously cleaned and sterilized equipment were submitted to contamination by spores. The experimental group was submitted to cleaning and sterilization by EtO and test of sterility by filtration through a 0.45 µm membrane. The membranes were cultivated and Gram and Wirtz-Conklin staining were carried out in positive results for spore visualization. An efficacy of 99.99999881 percent of the process of sterilization by EtO was confirmed, with a probability of survival of 1.19 x 10-8. Under the development conditions of the experiment, the efficacy of the sterilization of ED by EtO was confirmed.