A comparison of transcutaneous bilirubinometers: SpectRx BiliCheck versus Minolta AirShields.

Noch nicht übersetzt Noch nicht übersetzt
Autoren
Kategorie Primary study
ZeitungArchives of disease in childhood. Fetal and neonatal edition
Year 2002
BACKGROUND: Two devices are available for making transcutaneous estimates of serum bilirubin (SBR): the Minolta AirShields JM102 and the new SpectRx BiliCheck. OBJECTIVES: (a) To measure how well the readings produced by these devices agree with SBR measured in the laboratory; (b) to estimate for each device, the proportion of infants with clinical jaundice who would require blood sampling if the device was used as a screening tool to detect infants with SBR > or = 250 micromol/l. DESIGN: Prospective cohort study of jaundiced infants who required SBR at < or = 20 days of postnatal age. Those who had received phototherapy or exchange transfusion were excluded. SETTING: Tertiary neonatal service in South-East Scotland. INTERVENTIONS: Within 30 minutes of SBR sampling, transcutaneous bilirubinometry was performed using one Minolta and two SpectRx devices (designated A and B). RESULTS: Sixty-four neonates were enrolled, 19 of which were preterm (31-35 weeks). The 95% confidence intervals of a device reading corresponding to SBR were +/- 66.7, +/- 67.9, and +/- 66.4 micromol/l respectively. Using the devices to identify all SBR > or = 250 micromol/l would reduce SBR sampling by 23%, 16%, and 20% respectively. CONCLUSIONS: Given that SBR levels range from 50 to 400 micromol/l in jaundiced infants, the 95% confidence intervals of the devices are wide at +/- 67 micromol/l. The SpectRx can be used as a screening tool for hyperbilirubinaemia but there is no advantage in using it over the Minolta.
Epistemonikos ID: 143a0cbf607a891424b3a31ed8a7e51110a88649
First added on: Dec 23, 2016