Evaluation of flow cytometric counting procedure for canine reticulocytes by use of thiazole orange.

Ainda não traduzido Ainda não traduzido
Autores
Categoria Primary study
RevistaAmerican journal of veterinary research
Year 1991
An automated reticulocyte counting method that used a flow cytometer and the nucleic acid staining dye, thiazole orange, was developed. Anticoagulated (EDTA) blood specimens were suitable for flow cytometric reticulocyte counting when stored at 4 C for 96 hours after collection. Thiazole orange-stained samples were stable for 5.5 hours after staining when stored capped at 20 C and protected from light. Flow cytometric and manual microscopic reticulocyte counts were compared for counts in the 0.27 to 5.32% range (as determined by flow cytometry) and 0.10 to 4.90% range (as determined by 1 technician). Although the results of flow cytometric analysis generally correlated well (r = 0.821) with manual counts, there was poor correlation between the procedures for counts less than or equal to 2.0% (r less than or equal to 0.272). Linearity of flow cytometric counts over the range 0.27 to 14.46% was excellent (r = 0.999). Within-run precision of flow cytometric counts (% coefficient of variation [cv] = 3 to 5) was superior to manual microscopic counts obtained by one technician (% cv = 19 to 23) and to manual microscopic counts, which were an average of counts done by 3 technicians (% cv = 8 to 18). Comparable flow cytometric counts were obtained by counting 50,000 or 100,000 blood cells in the flow cytometer.