Comparison of the Biofire FilmArray RP, Genmark eSensor RVP, Luminex xTAG RVPv1, and Luminex xTAG RVP fast multiplex assays for detection of respiratory viruses.

Ainda não traduzido Ainda não traduzido
Autores
Categoria Primary study
RevistaJournal of clinical microbiology
Year 2013
There are several U.S. FDA-cleared molecular respiratory virus panels available today, each with advantages and disadvantages. This study compares four multiplex panels, the BioFire Diagnostics FilmArray RP (respiratory panel), the GenMark Dx eSensor RVP (respiratory viral panel), the Luminex xTAG RVPv1, and the Luminex xTAG RVP fast. Three hundred specimens (200 retrospective and 100 consecutive) were tested using all four platforms to determine performance characteristics. The overall sensitivity and specificity, respectively, and 95% confidence interval (CI; in parentheses) for each panel were as follows: FilmArray RP, 84.5% (79.2, 88.6) and 100% (96.2, 100); eSensor RVP, 98.3% (95.5, 99.5) and 99.2% (95.4, 100); xTAG RVPv1, 92.7% (88.5, 95.4) and 99.8% (96.0, 100); and xTAG RVP fast, 84.4% (78.5, 88.9) and 99.9% (96.1, 100). The sensitivity of each assay fluctuated by viral target, with the greatest discrepancies noted for adenovirus and influenza virus B detection. Hands-on time and time to result were recorded and ease of use was assessed to generate a complete profile of each assay.
Epistemonikos ID: 510bdbe4c5676d832e4c121af527bb17b6d9d1e9
First added on: Oct 14, 2018