Optical fiber dosimeter for real-time in-vivo dose monitoring during LDR brachytherapy.

Categoria Primary study
GiornaleBiomedical optics express
Year 2020
Loading references information
An optical fiber sensor for monitoring low dose radiation is presented. The sensor, based on radiation sensitive scintillation material, terbium doped gadolinium oxysulphide (Gd2O2S:Tb), is embedded in a cavity of 700µm diameter within a 1mm plastic optical fiber. The sensor is compared with the treatment planning system for repeatability, angular dependency, distance and accumulated radiation activity. The sensor demonstrates a high sensitivity of 152 photon counts/Gy with a temporal resolution of 0.1 seconds, with the largest repeatability error of 4.1%, to 0.361mCi of Iodine-125 the radioactive source most commonly used in LDR brachytherapy for treating prostate cancer.
Epistemonikos ID: 6ebb26e962de023caf922bdff383f8a50703af4f
First added on: Nov 08, 2024