Validation of portable noninvasive blood pressure monitoring devices: Comparisons with intra-arterial and sphygmomanometer measurements

Category Primary study
JournalAmerican Heart Journal
Year 1988
We have evaluated the accuracy of measurements provided by different types of portable automatic blood pressure monitoring devices by comparing them with intra-arterial and mercury sphygmomanometer measurements in 25 hospitalized patients. Systolic blood pressure values with portable devices that use auscultatory or oscillometric methods of measurement correlated significantly with intra-arterial values (r = 0.74 and 0.89; p < 0.001 for both); similarly, diastolic values correlated significantly (r = 0.86 and 0.81; p < 0.001 for both). Compared with intra-arterial measurements, there was a slight tendency for the portable devices to underestimate systolic blood pressure and overestimate diastolic blood pressure. Correlations between auscultatory or oscillometric measurements and sphygmomanometer measurements in these patients were also significant; moreover, the absolute blood pressure values obtained with the portable devices were almost identical to those with the sphygmomanometer. In a further group of 12 volunteers, auscultatory and oscillometric values correlated very closely with simultaneously measured sphygmomanometer values for both systolic (r = 0.88 and 0.98) and diastolic (r = 0.96 and 0.94) blood pressures. An auscultatory device that uses continuous ECG R wave gating for Korotkoff sounds was also found to be highly accurate. Thus we have found that automated portable devices that use either auscultatory or oscillometric methods of measurement provide reliable blood pressure values.
Epistemonikos ID: eb1ccc3a98dc3f3b412864c7e7dfcc993f8e19fd
First added on: Apr 18, 2026