Performance of third-generation confirmatory tests for detection of antibody to hepatitis C virus.

Authors
Category Primary study
JournalJournal of clinical microbiology
Year 1994
We investigated three immunoblot assays (RIBA 3.0 from Chiron, Matrix from Abbott Laboratories, and LiaTek III from Organon Teknika) for the detection of antibody to hepatitis C virus. RIBA 3.0 and Matrix require reactivity to two antigens and LiaTek III requires reactivity to one for a sample to be positive. We tested 80 samples that were positive in repeat enzyme immunoassays by supplemental tests. The results showed that 55, 46, and 28% were reactive by LiaTek III, RIBA 3.0, and Matrix, respectively; 54, 33, and 13% of the samples were indeterminate by Matrix, RIBA 3.0, and LiaTek III, respectively. There were 32, 21, and 16% nonreactive samples by LiaTek III, RIBA 3.0, and Matrix, respectively. Of the samples positive by RIBA 3.0, only 50 and 76% were reactive by Matrix and LiaTek III, respectively. A large number of samples that were indeterminate by RIBA 3.0 were positive by LiaTek III (52%). The core antigen was the most reactive antigen in all three tests (48 to 57%). The NS4 antigen in Matrix (20%) and LiaTek III (16%) was poorly reactive, although it performed better in RIBA 3.0 (45%). The NS5 and E2/NS1 antigens made minor contributions to reactivity. The combinations of the core, NS3, and NS4 antigens produced 77% of the RIBA 3.0 and 100% of the Matrix reactive samples. The results showed a poor correlation among the three tests.