A Novel 2-Hz Coronary Intravascular Lithotripsy Catheter for Calcified Coronary Artery Disease: One-Year Outcomes of the DISRUPT CAD Duo Trial

Authors
Category Primary study
JournalThe American journal of cardiology
Year 2026
Intravascular lithotripsy (IVL) modifies vascular calcification to enhance vessel compliance and optimize coronary stent deployment. A novel 2-Hz coronary IVL catheter was developed to improve procedural efficiency by shortening treatment time. Non-inferiority of the 2-Hz catheter for early safety and effectiveness has been previously reported, but there are no data on longer-term outcomes.The DISRUPT CAD Duo IDE was a multicenter, prospective, single arm study. Patients with de novo, severely calcified coronary lesions were treated with the 2-Hz IVL catheter. The primary safety endpoint was freedom from 30-day MACE (cardiac death, MI, or TVR) and the primary effectiveness outcome was procedural success (stent delivery with ≤30% residual stenosis, and no in-hospital MACE). Additional patient-level outcomes, including mortality, MI, revascularization, stent thrombosis, and anginal status were evaluated through 1-year follow-up.Of 145 enrolled patients, 143 were eligible for and 140 (97.9%) completed 1-year follow-up (non-eligible patients included: 1 non-cardiac death and 1 consent withdrawal). Through 1 year, 3 patients had MI, 2 of which were attributable to a target vessel; 5 patients required TVR, including 3 ischemia-driven TLR. The overall MACE rate was 5.6% (8/142) with no cardiac deaths or stent thrombosis. One non-cardiac death was reported. Compared with 14.5% of patients at baseline, 77.1% of patients reported having no angina symptoms at 1 year.The 2-Hz coronary IVL catheter, when used in patients with calcified coronary artery lesions, demonstrated sustained low rates of MACE alongside high rates of improvement in anginal symptoms through 1 year.
Epistemonikos ID: c83f05536d5fd9c099b06c0c5ab9133e63dac187
First added on: Sep 20, 2026