[A Comparative Study on the Damage Rate of Flow Rate Trigger and Pressure Trigger During the Process of Mechanical Ventilation and Spray].

Authors
Category Primary study
JournalZhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation
Year 2016
OBJECTIVE: To investigate the damage rate of the air flow sensor caused by the flow velocity and pressure in the process of mechanical ventilation and spray. METHODS: 12 wire type flow sensor were randomly divided into flow rate trigger group and pressure trigger group. Every time sprayed 15 minutes, then removed flow sensor, a total number of times gathered statistics when flow sensor failed, statistical analysis was performed using SPSS software. RESULTS: The flow rate trigger group was 410.83±29.8 times and was significantly less than that of the pressure trigger group of 955.67±84.54 times, the difference was statistically significant (P<0.05). CONCLUSIONS: The flow rate trigger during the process of mechanical ventilation and spray was significantly increased the damage rate of output flow sensor comparing with pressure trigger.
Epistemonikos ID: ad00a8d4ba98785ea83bb289aeb94e1ef79a4961
First added on: Sep 18, 2023