Modified ARCHITECT® serologic assays enable plasma-level performance from dried blood spot samples.

Category Primary study
JournalBioTechniques
Year 2022
Dried blood spots (DBSs) provide an alternative sample input for serologic testing. We evaluated DBSs for the ARCHITECT® hepatitis B surface antigen (HBsAg) NEXT, hepatitis B e-antigen (HBeAg), anti-hepatitis B core antigen (anti-HBc II), HIV antigen/antibody (Ag/Ab) Combo and AdviseDx SARS-CoV-2 IgG II assays. Assay performance with DBSs was assessed with or without assay modification and compared with on-market assay with plasma samples. DBS stability was also determined. HBsAg NEXT and HIV Ag/Ab Combo assays using DBSs showed sensitivity and specificity comparable to that of on-market assays. Modified HBeAg, anti-HBc II and SARS-CoV-2 IgG II DBS assays achieved performance comparable to on-market assays. Use of DBSs as input for high-throughput serologic assays is expected to have significant implications for improving population surveillance and increasing access to diagnostic testing.
Epistemonikos ID: 696b540b7ec28fd1a7db8087e36e318bb2068150
First added on: Oct 15, 2022