Comparison and Clinical Utility of the Portable Pressure Measuring Device for Garment Pressure Measurement on Hypertrophic Scar by Burn Injury During Compression Therapy

Category Primary study
Registry of Trialsclinicaltrials.gov
Year 2020
The investigators developed a portable pressure measuring device using silicon piezoresistive pressure sensors. As PicoPress® is the most accurate (i.e., lowest variation and error) manometric sensor for pressure measurement, the investigators used it to compare and examine the accuracy of the proposed device regarding in vitro pressure measurements. The purpose of this study was to determine the effectiveness of pressure garment therapy using proposed device with objective data obtained with a randomized within wound comparison. Pressure measurements were acquired through a readout circuit consisting of an analog-to-digital converter, a microprocessor, and a Bluetooth transmission module for wireless data transmission to an external device. The mean pressure values measured by the sensors were compared to those obtained from PicoPress®. This was a double-blinded, randomized, controlled trial of patients with hypertrophic scars. In the pressure monitoring group, garment pressures were monitored using the portable pressure measuring device, and the compression garment was adjusted so that the pressure was maintained at the therapeutic range of 15 - 25 mmHg. In the control group, non-surgical standard treatment of burn scars except for pressure monitoring was performed in the same manner.