Evaluation of a chemiluminescent enzyme immunoassay-based high-throughput SARS-CoV-2 antigen assay for the diagnosis of COVID-19: The VITROS® SARS-CoV-2 Antigen Test.

Category Primary study
JournalJournal of medical virology
Year 2021
A high-throughput, fully automated antigen detection test for SARS-CoV-2 is a viable alternative to reverse transcription polymerase chain reaction (RT-qPCR) for mass screening during outbreaks. In this study, we compared RT-qPCR for viral load and the VITROS® SARS-CoV-2 Antigen Test with reference to the results of the LUMIPULSE® SARS-CoV-2 Ag Test. Of 128 nasopharyngeal swab specimens taken from patients suspected of being infected with SARS-CoV-2, 49 were positive and 79 were negative according to RT-qPCR. Consistent dose-dependent detection with VITROS® assay was successfully achieved when using nasopharyngeal swab specimens with Ct values of ≤32.0, whereas the CLEIA-based the LUMIPULSE® assay was able to detect lower viral loads compared with the VITROS® assay. Our results show that the performance of the VITROS® assay was satisfactory for the diagnosis of contagious COVID-19 patients in the clinical setting. This article is protected by copyright. All rights reserved.