Comparative Evaluation of the Loop-Mediated Isothermal Amplification Assay for Detecting Pulmonary Tuberculosis

Category Systematic review
JournalAnnals of Laboratory Medicine
Year 2018
BACKGROUND: Early detection of tuberculosis (TB) is challenging in resource-poor settings because of limited accessibility to molecular diagnostics. The aim of this study was to evaluate the performance of the loop-mediated isothermal amplification kit (TB-LAMP) for TB diagnosis compared with conventional and molecular tests. METHODS: A total of 290 consecutive sputum samples were collected from May till September, 2015. All samples were processed using the N-Acetyl-L-cysteine (NALC) NaOH method and tested by smear microscopy, solid and liquid culture, real-time PCR, and TB-LAMP. RESULTS: The sensitivity of TB-LAMP for smear-positive and smear-negative samples with culture positivity was 92.0% and 58.8%, respectively. TB-LAMP was positive in 14.9% of TB culture-negative samples; however, all those samples were also positive by real-time PCR. In addition, none of the samples positive for nontuberculous mycobacteria by culture were positive by TB-LAMP. The overall agreement between TB-LAMP and real-time PCR was good; however, the concordance rate was significantly lower for real-time PCR positive samples with Ct values of 30–35. CONCLUSIONS: TB-LAMP could replace smear microscopy and increase TB diagnostic capacity when Xpert MTB/RIF is not feasible because of poor infrastructure.
Epistemonikos ID: 1c9d1669b602dd1c1b34a42e4a1a3d25fabcde84
First added on: Nov 28, 2024